PDF Download
CCHT 2025 WATER EXAM QUESTIONS
Actual Qs and Ans - Expert-Verified Explanation -Guaranteed passing score -18 Questions and Answers
-Format: Multiple-choice / Flashcard
Question 1: parts per million (ppm)
Answer:
how water hardness is measured
Question 2: What if Chlorine is noted in Post RO
Answer:
DO NOT INITIATE PT TX
Question 3: CARBON TANK 2 (POLISHER)
Answer:
IF TANK SHOWS RESULTS ABOVE THE ACTION LEVEL IMMEDIATELY STOP TREATMENTS
Question 4: RPC ultra low total chlorine strips
Answer:
Chlorine in water after carbon tank & post RO
Question 5: how long are rpc strips
Answer:
SAMPLE WATER FOR 60 SECONDS
Question 6: Bypass
Answer:
you will put the patient in this if carbon tank 2 test positive for chlorine.
Question 7: when should chlorine be tested?
Answer:
PRIOR TO TREATMENTS IN THE MORNING AND EVERY 3-4 HOURS WHILE PATIENTS ARE
DIALYZING
Question 8: Chlorine breakthrough
Answer:
OCCURS POST CARBON TANK 1 (WORKER), IMMEDIATE ADDITIONAL TESTING SHOULD TAKE
PLACE AFTER CARBON TANK 2 (POLISHER)
Question 9: Chlorine Testing
Answer:
REVERSE OSMOSIS (RO) MACHINES RUNS 15 MINUTES BEFORE PERFORMING ANY TOTAL
CHLORINE TESTING
Question 10: First sample
Answer:
SHOULD BE TAKEN FROM POST CARBON TANK 1 (WORKER )
Question 11: HOW OFTEN WATER SOFTNER TEST DONE
Answer:
THERE ARE 2 WATER HARDNESS TEST DONE DAILY, UNDER NORMAL OPERATING
CONDITIONS
Question 12: water hardness
Answer:
WATER EXISTING THE WATER SOFTENR
Question 13: What is the maximum amount of urine that can be collected in a urine cup?
Answer:
100ml
Question 14: RPC
Answer:
strips used for chlorine test
Question 15: acceptable limits for water hardness post softner
Answer:
20PPM Question 16: Total chlorine levels that are patient safe and also safe to document
Answer:
0.00-0.09
Question 17: Who must approve to resume normal operation?
Answer:
medical director