PDF Download FREE HEALTH & SOCIAL CARE AND STUDY GAMES ABOUT
CH. 13 AND 14 EXAM QUESTIONS
Actual Qs and Ans Expert-Verified Explanation
This Exam contains:
-Guarantee passing score -20 Questions and Answers -format set of multiple-choice -Expert-Verified Explanation Question 1: The higher the bioburden on an object:
Answer:
the more difficult it will be to sterilize Question 2: Which of the following statements about immediate use steam sterilization is true?
Answer:
It should be used only when there is not time to process items using the wrapped method Question 3: The process by which instruments are steam sterilized for immediate use is called:
Answer:
immediate use steam sterilization Question 4: Both the Association of periOperative Registered Nurses and The Joint
Commission recommend that:
Answer:
the use of immediate use steam sterilization be minimized or decreased
Question 5: When loading a steam sterilizer, basins should be:
Answer:
placed on edge Question 6: When transporting items that have been immediate use steam sterilized, it is
required that:
Answer:
they be transported in such a manner that reduces the potential for contamination Question 7: Implantable devices:
Answer:
should not be immediate use steam sterilization sterilized unless there is a tracking system in place to trace the item to a patient
Question 8: The coolest place in a steam sterilizer is the
Answer:
thermostatic trap Question 9: Before an item can be placed in a liquid chemical sterile processing system they must be
Answer:
cleaned
Question 10: The most common reason for steam sterilization failure is
Answer:
lack of steam contact with the instrument Question 11: The best way to transport items that have been processed using immediate use seam sterilization at the point of use
Answer:
in a rigid container designed for immedite use steam steriliation Question 12: How frequently should a sterilizer's strainer be removed and cleaned?
Answer:
Daily
Question 13: Central Service technicians need to understand the anatomy of a steam sterilizer to
Answer:
understand how the sterilizer operates Question 14: Items to be disinfected or sterilized at the point of use must be:
Answer:
properly cleaned per Manufacturer's Instructions for Use Question 15: Factors that cause sterilant contact failure with the instrument are:
Answer:
crowded loads, clogged drain strainer Question 16: Immediate use steam sterilization documentation should include:
Answer:
all of the above
Question 17: Three of the main phases of a terminal steam sterilizer cycle are
Answer:
conditioning, exposure and exhaust
Question 18: One of the most frequent causes of a clogged drain screen is
Answer:
tape
Question 19: The weakest part of a steam sterilizer is the
Answer:
door Question 20: When water is seen on the outside of a pack after sterilization the pack is considered safe to use if all other packs in the load are dry
Answer:
False