IBHRE Prep Exam Latest Update - Actual Exam 150 Questions and 100% Verified Correct Answers Guaranteed A+
A patient on flecainide should? - CORRECT ANSWER: Have their threshold checked
A PM that paces and sense only in the ventricle and is inhibited by spontaneous
ventricular activity is designated? - CORRECT ANSWER: VVI
A SSI PM is one that? - CORRECT ANSWER: senses and pace in either the atrium or
ventricle and inhibits
All of the following functions are programmable in both VVI and DDD PM's, except? -
CORRECT ANSWER: (X) AV interval duration
Sensitivity Refractory Output
An advantage of bipolar leads is? - CORRECT ANSWER: They are less susceptible to EMI
As pulse duration is shortened below .3mse., stimulation threshold increases as
described by all of the following, except? - CORRECT ANSWER: Energy
(x) Charge Volts Current
At the time of implant we want a threshold? - CORRECT ANSWER: at .5ms to be less than 1 volt for ventricle
Cardiac stimulation thresholds are expressed in terms of? - CORRECT ANSWER:
Voltage, current, and energy.
Characteristic findings in patients with the PM syndrome include any of the following
except? - CORRECT ANSWER: Febrile signs of PM's infection
Current PM's employ which of the following chemical batteries? - CORRECT ANSWER: Lithium-iodide
For a patient with evidence of sinus node dysfunction and intermittent heart block, which
of the following pacemakers would be inappropriate? - CORRECT ANSWER: DDDR
(x) AAI VVI
VVIR 1 / 3
for a wire fracture the resistance? - CORRECT ANSWER: increases
For voltage of 5V, resistance of 500 ohms and pulse width of .5ms, calculate the energy
delivered? - CORRECT ANSWER: 25 microjoules
Hysterisis is? - CORRECT ANSWER: The interval the PM waits after a sensed beat
before pacing
If a unipolar lead wire has an insulation break one would expect the resistance to? -
CORRECT ANSWER: decrease
If an external PM is set to a sensitivity setting of 5mV and some R-waves go unsensed,
then you should do the following - CORRECT ANSWER: Increase sensing by lowering
sensitivity to 2.5mV.
if the output vlotage of the pacemaker is programmed from 5 volts to 2.5 volts, the
energy delivered to the heart is? - CORRECT ANSWER: quartered
If the PM output voltage is 5v and the measured lead resistance is 330 ohms, then the current that flows out of the pacemaker into the heart is? - CORRECT ANSWER: 15.15 I=V/R=5V/.33 k ohm
In a bipolar pacing ventricular PM? - CORRECT ANSWER: Both the anode and the
cathode are in the heart.
PM programmability allows which of the following? - CORRECT ANSWER: Modifying
PM function in response to changing pt. status, diagnosing PM ECG's, prolongation of battery life.
Setting a low pacing rate on a demand ventricular PM may have all of the following
benefits except? - CORRECT ANSWER: Allowing a lower output setting.
The acute threshold peaking is due to? - CORRECT ANSWER: Trauma and
inflammation
The formula for calculating rate? - CORRECT ANSWER: 60,000 ms/min - time period in ms.
The slew rate is? - CORRECT ANSWER: Change in voltage divided by time or the
slope of the EGM
The specific intracardiac event sensed by PM's is termed? - CORRECT ANSWER:
Intrinsic deflection
- / 3
The strength duration curve is dependent on the? - CORRECT ANSWER: Cell's
membrane capacitance, intracellular and extracellular resistance and threshold voltage.
The two parameters that are important for ventricular sensing and should be measured
are? - CORRECT ANSWER: R-wave amplitude and slew rate
Typical acceptable P wave amplitude values for an acute atrial lead is? - CORRECT
ANSWER: greater than 2mV
Typical acceptable R-wave amplitude values for an acute ventricular lead is? -
CORRECT ANSWER: 7-15mV
Typical lead resistance range measured in ohms is? - CORRECT ANSWER: 300-2000
ohms
Voltage is? - CORRECT ANSWER: The electromotive force pushing on electrons
What is the initial voltage of a lithium battery? - CORRECT ANSWER: 2.78V
When connecting a unipolar PM to the lead in the patient? - CORRECT ANSWER:
place the PM in the pocket to check pacing
Which of the following affect the longevity of the PM? - CORRECT ANSWER: output
voltage, resistance of lead and amount % pacing
Which of the following combinations of variables interact to determine stimulation
threshold? - CORRECT ANSWER: Voltage and pulse width
Which of the following factors does not contribute to development of the PM syndrome
w/ventricular pacing? - CORRECT ANSWER: Hysterisis
Which of the following is likely to happen during the lead maturation process? -
CORRECT ANSWER: Encapsulation of electrode with fibrin tissue
Which of the following is not an advantage of DD PM? - CORRECT ANSWER: Two
leads are required
Which of the following is/are NOT factor (s) which influence PM longevity? - CORRECT
ANSWER: Output voltage
lead resistance pulse duration (X) blanking
Which of the following statements is not true of rate-responsive ventricular PM's that
use a sensor other than the atrium? - CORRECT ANSWER: They maintain constant AV
synchrony
- / 3