ASSE 6030 Medical Gas Systems Verifier Practice Exam Questions And Correct Answers (Verified Answers) Plus Rationales 2026 Q&A | Instant Download Pdf
- Which organization establishes the qualifications for medical gas verifiers
under the ASSE 6000 series?
A. NFPA
B. ANSI
- ASSE International
D. OSHA
Rationale: ASSE International develops and maintains the ASSE 6000 series
professional qualification standards for medical gas personnel.
- / 4
2. The ASSE 6030 verifier is responsible for:
- Designing the medical gas system
- Verifying the installation and performance of the system
- Maintaining the system daily
- Certifying the installer
Rationale: The verifier ensures the system is installed per NFPA 99 and
performs functional verification testing.
- What is the primary standard governing medical gas systems in healthcare
facilities?
A. ASME B31.1
B. NFPA 99
C. NFPA 101
D. ASSE 6000
Rationale: NFPA 99 outlines requirements for the design, installation, and
verification of medical gas and vacuum systems.
4. According to NFPA 99, the verifier shall be independent of:
- The facility’s maintenance team only
- The manufacturer only
- The installer and manufacturer
- The AHJ 2 / 4
Rationale: The verifier must be independent to ensure unbiased
verification results.
5. The final verification report must be submitted to:
- The installing contractor
- The facility’s responsible authority or owner
- The gas supplier
- The design engineer
Rationale: The owner or responsible authority receives the official
verification documentation.
- Which document must the verifier review before beginning verification
- Maintenance schedule
- Equipment warranty
- Installer’s certification documents and as-built drawings
- Owner’s manual
testing?
Rationale: Verification is based on installed conditions and system
drawings to confirm compliance.
- What is the required purity for medical oxygen in the pipeline system?
A. 90% 3 / 4
B. 93%
- 99.0% minimum
D. 99.9%
Rationale: Medical oxygen must meet the U.S. Pharmacopeia (USP)
requirement of not less than 99.0% purity.
- Which of the following gases is classified as nonflammable and oxidizing?
- Nitrogen
- Oxygen
- Carbon dioxide
- Nitrous oxide
Rationale: Oxygen supports combustion and is classified as an oxidizer,
though not flammable.
- Vacuum systems for patient use must maintain a minimum flow rate per
NFPA 99 of:
- 2.5 scfm per station
- 3.5 scfm per station
- 5.0 scfm per station
- 10 scfm per station
Rationale: The required flow rate for patient vacuum outlets is 3.5 scfm at
12 inHg minimum.
- / 4